INVIMA Medical Device Registration in Colombia (2027 Guide)
What INVIMA requires to import medical devices into Colombia in 2027: risk classes, timelines, documents, the CCAA, registration checks and the new regime ahead.

Key Takeaways
- Selling an imported medical device in Colombia requires an INVIMA sanitary registration (or a marketing permit for controlled biomedical equipment) under Decree 4725 of 2005.
- Class I and IIa devices get automatic registration in 2 business days; Class IIb and III go through review of up to 90 business days. Registrations last 10 years.
- The importer needs a CCAA certificate (Resolution 4002 of 2007), valid for 5 years per site.
- Before buying, check the number, holder, model and validity in INVIMA's registration lookup.
- A new unified device regime is in draft, with an 18-month transition once published.
Short answer: to import and sell a medical device in Colombia you need a sanitary registration (registro sanitario) from INVIMA, the national food and drug regulator, issued to a holder that answers for the product in the country. The rules are set by Decree 4725 of 2005: Class I and IIa devices receive automatic registration within 2 business days, Class IIb and III devices go through technical review of up to 90 business days, and registrations are valid for 10 years.
This 2027 guide is written for importers, distributors and procurement teams handling respiratory equipment such as CPAP, BiPAP and oxygen concentrators. For the regional picture, see our medical device registration guide for Latin America.
What is an INVIMA registration and who needs one?
The registro sanitario is the public document through which INVIMA authorizes the production, import or sale of a medical device after checking its technical, legal and sanitary requirements. It is issued for a product, a manufacturer and a holder. For controlled-technology biomedical equipment in Class IIb and III, the equivalent authorization is a marketing permit (permiso de comercialización).
The registration holder answers to INVIMA: it files the application, keeps the registration current and reports adverse events. In practice this is usually the importer, a distributor or a regulatory firm based in Colombia, authorized by the manufacturer.
What are the risk classes and how long does registration take?
Decree 4725 sorts devices into four risk classes: I, IIa, IIb and III. The class sets the procedure and timeline:
| Risk class | Procedure | INVIMA timeline | Validity |
|---|---|---|---|
| I and IIa | Automatic sanitary registration | 2 business days from filing | 10 years |
| IIb and III | Registration with technical and legal review | 90 business days from filing | 10 years |
| Controlled biomedical equipment, IIb and III | Marketing permit | Per procedure | 10 years |
The applicant determines the class using the decree's rules, and INVIMA reviews it. As a reference point, the user manual for the OT series concentrators lists Class IIa under its CE marking; the Colombian class must still be confirmed under Decree 4725 for each model.
A caution on automatic registration: if INVIMA requests more information, the applicant has 90 business days to respond. Otherwise the registration is suspended, and it is cancelled if the issue is not fixed within three months.
What does INVIMA require for an imported device?
The legal requirements are the same for every class:
- Applicant name and payment of the procedure fee.
- Power of attorney, if a lawyer handles the filing.
- Proof of existence and legal representation of the manufacturer, importer or holder.
- A certificate of free sale from the country of origin, legalized or apostilled, with an official translation if not in Spanish.
- The manufacturer's authorization to the Colombian importer or holder.
Technical requirements vary by class:
- Quality system certification (GMP or CCAA), for all classes.
- Device description, technical studies and analytical tests, declaration of conformity, sterilization and disposal method, and shelf life where applicable.
- Original label artwork, importer sticker and inserts, in Spanish.
- List of applied standards and commercial history.
- Risk analysis and scientific information, from Class II up.
- Clinical studies and implant card, for IIb and III.
- For Class I and II biomedical equipment: operation and maintenance manuals in Spanish.
Devices from reference countries such as the European Union, the United States, Canada, Japan and Australia can submit homologated documents under Decree 3275 of 2009. Confirm the current list on INVIMA's medical devices and biomedical equipment page before building the file.
What is the CCAA certificate and who needs it?
The storage and conditioning capacity certificate (CCAA) is issued by INVIMA to the importer and confirms compliance with Resolution 4002 of 2007. It is a prerequisite for a sanitary registration and is valid for 5 years.
- Each site where product is imported, received, stored, conditioned or released needs its own certificate; moving requires a new one.
- Distributors that do not import do not need an INVIMA CCAA; territorial health authorities supervise them.
- The technical director is part of the CCAA, and any change must be notified to INVIMA.
For a distributor starting to import, the CCAA is often the critical path: no certified warehouse, no registration.
How do you verify an INVIMA registration?
Before buying, or before listing a product in a tender, check it in INVIMA's sanitary registration lookup. Review five items:
- Registration number, and that it is current, not expired, suspended or cancelled.
- Holder: the company answerable for the product.
- Manufacturer and country of origin.
- Models and references covered: one registration may cover some models and not others.
- Authorized importers, if you buy from a distributor that is not the holder.
Ask your supplier for the registration number of every quoted model and compare it with the lookup. If the number is missing or the model does not match, resolve it before issuing the purchase order.
What obligations apply after registration?
- Technovigilance: the national technovigilance program, regulated by Resolution 4816 of 2008, requires reporting of adverse events and incidents.
- Modifications: changes to a registration are filed with INVIMA; changes that affect safety or effectiveness cannot go through the automatic route.
- Renewal: if a renewal is filed on time and complete, the current authorization stays in effect until INVIMA decides.
What changes in 2027 with the new medical device regime?
The Ministry of Health is preparing a unified regime intended to replace the current frameworks, including Decree 4725 of 2005 and Decree 3770 of 2004. As of mid-2026 it was still a draft: it was notified to the WTO on 18 May 2026, with comments open until 17 July 2026, and it still needs legal review and the President's signature.
According to the draft discussed by INVIMA:
- Authorizations would have indefinite validity instead of today's 10 years, conditional on risk.
- A unique device identifier (UDI) would be required from the initial application.
- Reliance mechanisms would recognize reviews by agencies such as FDA, EMA or ANVISA, without an automatic fast track.
- In vitro diagnostic reagents would join the device regime.
- Once published, there would be an 18-month transition.
Colombia has been an IMDRF affiliate member since September 2025, which pushes alignment with international guidance. Until the new decree is published, Decree 4725 remains the applicable rule.
How should respiratory equipment importers prepare?
- Confirm the risk class of each model: CPAP, BiPAP, oxygen concentrators and masks can fall into different classes.
- Ask the manufacturer early for the apostilled certificate of free sale, the importer authorization and Spanish manuals.
- Secure the CCAA for your warehouse before filing.
- Plan renewals months ahead, not at expiry.
SysMed USA supplies the technical documentation for the devices it distributes to support your holder's file. See our Colombia page or request a quote.
References
Frequently Asked Questions
How long does INVIMA registration take for a medical device?
It depends on the risk class. Class I and IIa devices get automatic sanitary registration within 2 business days of filing. Class IIb and III devices go through technical and legal review of up to 90 business days.
The clock runs on a complete file. Before that, the importer needs a CCAA certificate for its warehouse and manufacturer documents such as an apostilled certificate of free sale.
How long is an INVIMA registration valid?
Sanitary registrations and marketing permits for medical devices are valid for 10 years from the date of the administrative act. If a renewal is filed on time and complete, the current authorization stays in effect until INVIMA decides.
The draft of Colombia's new device regime proposes indefinite validity conditional on risk, but until it is published, Decree 4725 of 2005 applies.
How do I check whether a device has an INVIMA registration?
Use the sanitary registration lookup on INVIMA's website (invima.gov.co/consulta-registros-sanitarios). Search by registration number or product and confirm it is current, who the holder is, the manufacturer, the models covered and the authorized importers.
Ask your supplier for the registration number of every quoted model and compare it before issuing the purchase order.